ایرادات ریکو براساس کد می باشد شرکت کپی ایران در نظر داردتمامی سرویس منالها و رفع ایرادادت را در اختیار شما عزیزان قرار دهد

 

 

    1.  Exposure lamp error
      The standard white level was not detected properly when scanning the white platecopyir.com    • Exposure lamp defective • Lamp stabilizer defective • Exposure lamp connector defective • Standard white plate dirty • Scanner mirror or scanner lens out of position or dirty • SBU defective
      120    Scanner home position error 1
      The scanner home position sensor does not detect the on condition during initialization or copyingcopyir.com    • Scanner drive motor defective • Scanner motor defective • Harness between SBCU and scanner drive motor disconnected • Harness between SBCU and scanner drive motor power source disconnected • Scanner HP sensor defective • Harness between SBCU and HP sensor disconnected • Scanner wire, timing belt, pulley, or carriage defective
      121    Scanner home position error 2
      The scanner home position sensor does not detect the off condition during initializationcopyir.com    • SIB or scanner drive motor defective • Scanner motor defective • Harness between SBCU and scanner drive motor disconnected • Harness between SBCU and scanner drive motor power source disconnected • Scanner HP sensor defective • Harness between SBCU and HP sensor disconnected • Scanner wire, timing belt, pulley, or carriage defective
      122    Scanner HP Sensor œ Error 1
      The HP sensor remains on while the carriage is returning to the home positioncopyir.com    • SBCU, scanner motor drive board defective • Scanner motor defective • Harness between the SBCU, scanner motor drive board and scanner motor disconnected • HP sensor defective • Harness between SBCU and HP sensor disconnectedcopyir.com • Scanner wire, timing belt, pulley, or carriage installation incorrect
      123    Scanner HP sensor œ Error 2
      The HP sensor does not switch on after the carriage has returned to the home positioncopyir.com    • SBCU, scanner motor drive board defective • Scanner motor defective • Harness between the SBCU, scanner motor drive board and scanner motor disconnected • HP sensor defective • Harness between SBCU and HP sensor disconnectedcopyir.com • Scanner wire, timing belt, pulley, or carriage installation incorrect
      144    SBU connection error
      The IPU does not detect the SBU connection signalcopyir.com    Cable between SBU and IPU board disconnectedcopyir.com
      192    Automatic SBU adjustment error
      An error is detected during automatic SBU adjustment (SP4-428)    • SBU defective • SBCU board defective • Exposure lamp stabilizer defective • Exposure lamp defective • Dirty white plate
      193    Image transfer error
      The IPU board does not finish within 1 minute after the image data has been transferred to the controller boardcopyir.com    • IPU board defective • Controller board defective • Video controller defective
      195    DFGATE assert error
      The DFGATE signal does not assert within 30 seconds after the original has been scannedcopyir.com    • ADF interface cable defective • SBCU board defective • Mismatched firmware between the SBCU board and ADF
      196    DFGATE negate error
      The DFGATE signal does not negate within 1 minute after the DFGATE has been assertedcopyir.com    • ADF interface cable defective • SBCU board defective • Mismatched firmware between the SBCU board and ADF
      197    DFGATE error
      The DFGEATE signal has already been asserted at the original scancopyir.com    • ADF interface cable defective • SBCU board defective • Mismatched firmware between the SBCU board and ADF
      198    Memory address error
      The IPU board does not receive the memory address from the controller boardcopyir.com    • Mismatched firmware between the SBCU board and controller board • Controller defective • SBCU defective • IPU board defective
      199    DF scanning finish error
      The original does not finish scanning with in 1 minute    • ADF interface cable defective • SBCU board defective • Mismatched firmware between the SBCU board and ADF
      302    Charge roller current leak
      A charge roller current leak signal is detectedcopyir.com    • Charge roller damaged • Charge high voltage supply board defective • Poor connection of the PCU
      320    Polygon motor error
      The polygon motor does not reach its operating speed within 10 seconds after the polygon motor on signal, or the lock signal is not detected for more than 200 ms continuously during operationcopyir.com    • Polygon mirror motor defective • Poor connection between the polygon mirror motor driver and the SBCU board • SBCU board defective
      322    Laser synchronization error
      The main scan synchronization detector board cannot detect the laser synchronization signal for more than 10 consecutive 50 ms intervalscopyir.com    • Poor connection between the laser synchronization detector board and the SBCU board • Laser synchronization detector board out of position • Laser synchronization detector board defective • SBCU board defective • LD unit defective
      323    LD drive current over
      The LD drive board applies more than 100 mA to the LDcopyir.com    • LD unit defective (not enough power, due to aging) • Poor connection between the LD unit and the SBCU board • SBCU board defective
      350    ID sensor calibration - Error 1
      One of the following conditions occurred when the ID sensor pattern was calibrated during printing: • Vsp > 2copyir.com5V • Vsg < 2copyir.com5V • Vsp = 0V • Vsg = 0V    • ID sensor defective or dirty • ID sensor harness disconnected or connector is damaged • SBCU defective • Scanning system or image creation system malfunction • High voltage power supply board (power pack) defective
      351    ID sensor calibration œ Error 2
      The following conditions occurred simultaneously when the ID sensor pattern was calibrated during printing: • Vsg = 5V • PWM = 0 (LED current drop)    • ID sensor dirty or defective • ID sensor harness disconnected, or connector damaged • SBCU board defective • High voltage power supply board (power pack) defective
      352    ID sensor calibration œ Error 3
      During printing the 2copyir.com5V value for edge detection of the ID sensor pattern could not be detected after 800 mscopyir.com    • ID sensor dirty or defective • ID sensor harness disconnected, or connector damaged • SBCU defective • High voltage power supply board (power pack) defective
      353    ID sensor adjustment Error 1
      Error occurred during automatic adjustment of Vsg: • Vsg output did not attain 4V, even with PWM = 255 (maximum current for LED) • Vsg output was greater than 4V, even with PWM=0 (no current for the LED)    • ID sensor dirty or defective • ID sensor harness disconnected, or connector damaged • SBCU defective • High voltage power supply board (power pack) defective • Scanning system or image creation system malfunction
      354    ID Sensor Adjustment Error 2
      Error occurred during automatic adjustment of Vsgcopyir.com Vsg could not be adjusted to 4copyir.com0V±0copyir.com2V within 50 ms even after 20 attemptscopyir.com    • ID sensor dirty or defective • ID sensor harness disconnected, or connector damaged • SBCU defective • High voltage power supply board (power pack) defective • Scanning system or image creation system malfunction
      355    ID sensor error
      For details about the cause of the problem, please refer to SC350~354 abovecopyir.com    • ID sensor dirty or defective • ID sensor harness disconnected, or connector damaged • SBCU board defective • High voltage power supply board (power pack) defective • Scanning system or image creation system malfunction
      389    TD sensor error
      TD sensor output was less than 0copyir.com5V, or more than 0copyir.com5V 10 times in successioncopyir.com If the fax unit is installed, this SC is issued immediatelycopyir.com If the fax unit is not installed, this SC is issued after the prescribed number of copies has printedcopyir.com    • TD sensor defective • TD sensor connector damagedcopyir.com
      390    TD sensor error
      The TD sensor outputs less than 0copyir.com5V or more than 4copyir.com0V 10 times consecutively during copyingcopyir.com    • TD sensor abnormal • Poor connection of the PCU
      391    Development bias leak
      A development bias leak signal is detectedcopyir.com    • Poor connection of the PCU • High voltage supply board defective
      392    TD sensor initial setting error
      TD sensor output voltage falls out of the adjustment range (2copyir.com0 ±<> 0copyir.com2 V) after the TD sensor initial setting has been finishedcopyir.com    • Someone forgot to remove the toner seal of the PCU • ID sensor defective • TD sensor abnormal • Drum does not turn • Development roller does not turn • Poor connection of the PCU
      399    Illegal toner bottle (South Korea only)
      The toner bottle installed is not intended for use with this machinecopyir.com    • Install the correct type of toner bottlecopyir.com
      401    Transfer roller leak error 1
      A transfer roller current leak signal is detectedcopyir.com The current feedback signal for the transfer roller is not detectedcopyir.com    • High voltage supply board defective • Poor connection of the PCU • Transfer/separation unit set incorrectly • Transfer roller damaged
      402    Transfer roller leak error 2
      A transfer roller current leak signal is detectedcopyir.com The current feedback signal for the transfer roller is not detectedcopyir.com    • High voltage supply board defective • Poor connection of the PCU • Transfer/separation unit set incorrectly • Transfer roller damaged
      411    Separation bias leak error
      A separation bias leak signal is detectedcopyir.com    • High voltage supply board defective • Poor connection of the PCU • Discharge plate defective
      490    Toner supply motor leak error
      Over 1 A supplied to the toner supply motor for longer than 200 ms    Toner supply motor defective
      500    Main motor lock
      A main motor lock signal is not detected for more than 500 ms after the main motor starts to rotate, or the lock signal is not detected for more than 500 ms during rotation after the last signalcopyir.com    • Too much load on the drive mechanism • Main motor defective
      501    1st paper tray lift motor malfunction    • Paper lift sensor defective • Tray lift motor defective • Too much load on the drive mechanism • Poor tray lift motor connection
      502    2nd paper tray lift motor malfunction    • Paper lift sensor defective • Tray lift motor defective • Too much load on the drive mechanism • Poor tray lift motor connection
      503    3rd paper tray lift motor malfunction (optional paper tray unit)    • Paper lift sensor defective • Tray lift motor defective • Too much load on the drive mechanism • Poor tray lift motor connection
      504    4th paper tray lift motor malfunction (optional paper tray unit)    • Paper lift sensor defective • Tray lift motor defective • Too much load on the drive mechanism • Poor tray lift motor connection
      506    Paper tray motor lock (optional paper tray unit)
      A motor lock signal is not detected for more than 1copyir.com5 s or the lock signal is not detected for more than 1copyir.com0 s during rotationcopyir.com    • Paper tray unit motor defective • Too much load on the drive mechanism
      508    Rear fence drive error (optional LCT)
      The return position sensor is not activated after the rear fence drive motor has been on to lower the tandem tray for 8 secondscopyir.com    • Rear fence motor defective • Return position sensor defective • Too much load on the drive mechanism
      509    Side fence drive error (optional LCT)
      The side fence positioning sensor is not activated for more 3 seconds when the paper stack in the left tray is moved to the right traycopyir.com The side fence close sensor is not activated for more 3 seconds after moving the paper stack to the right traycopyir.com    • Side fence motor defective • Side fence position sensor defective • Side fence close sensor defective • Too much load on the drive mechanism
      510    LCT lower limit error
      The lower limit sensor does not activate within 8 seconds after the tray has been loweredcopyir.com    • Tray lift motor defective • Poor connection of the tray lift motor • Lower limit sensor defective • Too much load on the drive mechanism
      520    Paper tray error
      An error occurs (icopyir.come motor error, or sensor error, etc) for any paper traycopyir.com    • A defective motor • A defective sensor • Too much load on the drive mechanism
      541    Fusing thermistor open
      The fusing temperature detected by the thermistor was below 0 °C for 5 secondscopyir.com The fusing temperature does not rise +15 °C (center) or +12 °C (at the ends) five times within 2 minutes after the fusing lamps have been turned oncopyir.com    • Fusing thermistor defective or out of position • Fusing lamp open • Fusing thermostat open • Power supply board defective • Poor connection of the fusing unit
      542    Fusing temperature warm-up error
      The fusing temperature does not reach the standby temperature within 20 seconds after the main switch is turned oncopyir.com    • Fusing thermistor defective or out of position • Fusing lamp open • Fusing thermostat open • Power supply board defective • Poor connection of the fusing unit
      543    Fusing overheat error 1
      A fusing temperature of over 231°C is detected for 5 second by the fusing thermistorcopyir.com    • Fusing thermistor defective • Power supply board defective • SBCU board defective
      544    Fusing overheat error 2
      A fusing temperature of over 251°C is detected by the fusing temperature monitor circuit in the SBCU boardcopyir.com    • Fusing thermistor defective • Power supply board defective • SBCU board defective
      545    Fusing overheat error 3
      After warmup, the hot roller attained full operating temperature and maintained this temperature for 10 seccopyir.com without the hot roller rotatingcopyir.com    • Hot roller thermistor is out of position • Fusing lamp broken • Thermostat broken
      546    Unstable fusing temperature
      The fusing temperature varies 50°C or more within 1 second twice continuouslycopyir.com    • Thermistor defective • Poor connection of the fusing unit • Power supply unit defective
      547    Zero cross signal malfunction
      Zero cross signals are not detected within a certain periodcopyir.com    • Power supply board defective • SBCU board defective
      548    Fusing unit set error
      The machine does not detect the fusing unitcopyir.com    • Poor connection of the fusing unit • The fusing unit is not installed
      590    Exhaust fan motor error
      The CPU detects an exhaust fan lock signal for more than 5 secondscopyir.com    • Poor connection of the exhaust fan motor • Too much load on the motor drive
      611    Communication break error between SBCU and ADF
      The SBCU receives a break signal from the ADF main boardcopyir.com    • Serial line connecting SBCU and ADF unstable • External noise • ADF main board defective • SBCU board defective
      612    Communication command error between SBCU and ADF
      The SBCU sends a command to the ADF main board that it cannot executecopyir.com    • SBCU board defective • Download SBCU firmware again
      620    Communication timeout error between SBCU and finisher
      The SBCU cannot receive a response within 100 ms after 3 attempts after sending data to the finishercopyir.com    • Serial line connecting SBCU and finisher unstable • External noise • SBCU board and finisher main board connection defective or loose • Finisher main board defective • SBCU board defective
      621    Communication timeout error between SBCU and finisher
      A break (low) signal was received from the finishercopyir.com    • Serial line connecting SBCU and finisher unstable • External noise
      630    Communication failure with CSS (RSS)
      The communication from the copier was detected as abnormal at the CSS centercopyir.com This error occurs when the acknowledge signal from the LADP does not complete normallycopyir.com    • Occurred with an SC call, CC call, supply management call, user call, or CE callcopyir.com • Timeout while no response from the LADP, and abnormal signal on the RS-485 line between PI and LADPcopyir.com
      632    Counter device error 1
      After 3 attempts to send a data frame to the optional counter device via the serial communication line, no ACK signal was received within 100 mscopyir.com    • Serial line between the optional counter device, the relay board and copier control board is disconnected or damaged • Make sure that SP5113 is set to enable the optional counter devicecopyir.com
      633    Counter device error 2
      After 3 attempts to send a data frame to the optional counter device via the serial communication line, no ACK signal was received within 100 mscopyir.com    • Serial line between the optional counter device, the relay board and copier control board is disconnected or damaged • Make sure that SP5113 is set to enable the optional counter devicecopyir.com
      634    Counter device error 3
      A backup RAM error was returned by the counter devicecopyir.com    • Counter device control board defective • Backup battery of counter device defective
      635    Counter device error 4
      A backup battery error was returned by the counter devicecopyir.com    • Counter device control board defective • Backup battery of counter device defective
      640    SBCU control data transfer checksum error
      A sampling of control data sent from the SBCU to the controller reveals a checksum errorcopyir.com Only the logging count is performedcopyir.com    • Controller board defective • External noise • SBCU board defective
      641    SBCU control data transfer abnormal
      A sampling of the control data sent from the SBCU reveals an abnormalitycopyir.com    • Controller board defective • External noise • SBCU board defective
      650    Communication timeout error between SBCU and duplex unit
      The SBCU cannot receive a response within 1 second from the duplex unitcopyir.com    • Serial line connecting SBCU and duplex unit unstable • External noise • SBCU board and duplex main board connection defective or loose • Duplex main board defective • SBCU board defective
      670    Engine response error
      After powering on the machine, a response is not received from the engine within 30 secondscopyir.com    • SBCU installed incorrectly • SBCU defective • Controller board defective
      672    Controller-to-operation panel communication error at startup
      After powering on the machine, the communication circuit between the controller and the operation panel is not opened, or communication with controller is interrupted after a normal startupcopyir.com    • Controller stall • Controller board installed incorrectly • Controller board defective • Operation panel connector loose or defective
      720    Finisher jogger motor error (500-sheet finisher)
      The finisher jogger Hcopyir.comP sensor remains de-activated for a certain time when returning to home positioncopyir.com The finisher jogger Hcopyir.comP sensor remains activated for a certain time when moving away from home positioncopyir.com    • Jogger Hcopyir.comP sensor defective • Jogger motor defective
      722    Finisher jogger motor error (1000-sheet finisher)
      The finisher jogger Hcopyir.comP sensor remains de-activated for a certain time when returning to home positioncopyir.com The finisher jogger Hcopyir.comP sensor remains activated for a certain time when moving away from home positioncopyir.com    • Jogger Hcopyir.comP sensor defective • Jogger motor defective
      724    Finisher staple hammer motor error (1000-sheet finisher)
      Stapling does not finish within 600 ms after the staple hammer motor turned oncopyir.com    • Staple jam • Stapler overload caused by trying to staple too many sheets • Staple hammer motor defective
      725    Finisher stack feed-out motor error (1000-sheet finisher)
      The stack feed-out belt Hcopyir.comP sensor does not activate within a certain time after the stack feed-out motor turned oncopyir.com    • Stack feed-out Hcopyir.comP sensor defective • Stack feed-out motor defective
      726    Finisher lift motor error (1000-sheet finisher)
      The stack height sensor does not activate within a certain time after the shift tray lift motor turned oncopyir.com    • Shift tray lift motor defective • Stack height sensor defective
      727    Finisher staple hammer motor error (500-sheet finisher)
      Stapling does not finish within a certain time after staple hammer motor turned oncopyir.com    • Staple jam • Stapler overload caused by trying to staple too many sheets • Staple hammer motor defective
      728    Finisher paper stack height error (500sheet finisher)
      The stack height detection lever does not return to its home position before going to detect the stack heightcopyir.com    • Stack height lever solenoid defective • Stack height sensor defective • Lever sensor defective • Main control board defective
      730    Finisher stapler motor error (1000-sheet finisher)
      The stapler does not return to its home position within a certain time after the stapler motor turned oncopyir.com The stapler Hcopyir.comP sensor does not activate within a certain time after the stapler motor turned oncopyir.com    • Stapler motor defective • Stapler Hcopyir.comP sensor defective • Poor stapler motor connection
      731    Output tray motor error (500-sheet finisher) Exit guide plate motor error (1000-sheet finisher)
      The tray upper limit sensor does not activate within a certain time after the shift motor turned oncopyir.com (500-sheet finisher) The exit guide plate open sensor or exit guide plate HP sensor does not activate within a certain time after the exit guide plate motor turned oncopyir.com (1000-sheet finisher)    500-sheet finisher • Output tray motor defective • Tray upper limit sensor defective 1000-sheet finisher • Exit guide plate motor defective • Exit guide plate HP sensor defective • Exit guide plate open sensor defective
      732    Finisher shift motor error (1000-sheet finisher)
      Roller shift does not finish within a certain time after the shift motor turned oncopyir.com    • Shift motor defective • Shift tray HP sensor defective
      770    Shift tray shift motor error
      During a shift operation the sensor state did not change (off to on, or on to off) within 2copyir.com4 seccopyir.com    • Shift sensor defective • Shift motor defective
      791    Bridge communication error
      The SBCU cannot communicate with the bridge unit properly when the finisher is installedcopyir.com    • Poor connection between the main machine and bridge unit • SBCU board defective
      792    Finisher connection error
      The SBCU cannot communicate with the finisher properly when the bridge unit is installedcopyir.com    • Poor connection between the finisher and bridge unit • SBCU board defective
      793    Interchange communication error
      The SBCU cannot communicate with the interchange unit properly when the duplex unit is installedcopyir.com    • Poor connection between the interchange unit and main machine • SBCU board defective
      800    Startup without video output end error (K)
      Video transfer to the engine is started, but the engine did not issue a video transmission end command within the specified timecopyir.com    Controller board defective
      804    Startup without video input end (K)
      A video transmission was requested from the scanner, but the scanner did not issue a video transmission end command within the specified timecopyir.com    Controller board defective
      818    Watchdog error
      The CPU does not access the watchdog register within a certain timecopyir.com    • Controller board defective • Software malfunction œ download controller firmware again
      819    Kernel mismatch error
      Software bug    Download controller firmware again
      820    Self-Diagnostic Error: CPU
      The central processing unit returned an error during the self-diagnostic testcopyir.com    • Controller board defective • Download controller firmware again
      821    Self-Diagnostic Error: ASIC
      The ASIC returned an error during the self-diagnostic test because the ASIC and CPU timer interrupts were compared and determined to be out of rangecopyir.com    Controller board defective
      822    Self-Diagnostic Error: HDD
      The hard disk drive returned an error during the self-diagnostic testcopyir.com    • HDD defective • HDD connector defective • Controller board defective
      823    Self-diagnostic Error: NIB
      The network interface board returned an error during the self-diagnostic testcopyir.com    • Network interface board defective • Controller board defective
      824    Self-diagnostic Error: Resident NVRAM
      The resident non-volatile RAM returned an error during the self-diagnostic testcopyir.com    • Replace the resident NVRAM on the controller board • Replace the controller board
      825    Self-diagnostic Error: Optional NVRAM
      The resident non-volatile RAM returned an error during the self-diagnostic testcopyir.com    • Replace the optional NVRAM (user account enhancement unit) on the controller board • Replace the controller board
      826    Self-diagnostic Error: NVRAM/Optional NVRAM
      The NVRAM or optional NVRAM returned an error during the self-diagnostic testcopyir.com    Replace the NVRAM on the controller board
      827    Self-diagnostic Error: RAM
      The resident RAM returned a verify error during the self-diagnostic testcopyir.com    Download controller firmware again
      828    Self-diagnostic Error: ROM
      The resident read-only memory returned an error during the self-diagnostic testcopyir.com    • Controller board defective • Download controller firmware again
      829    Self-diagnostic Error: Optional RAM
      The optional RAM returned an error during the self-diagnostic testcopyir.com    • Replace the optional memory board • Controller board defective
      835    Self-Diagnostic Error: Parallel Interface
      Loopback test errorcopyir.com    • Loopback connector not detected • IEEE1284 connector defective • Controller board defective
      836    Self-diagnostic Error: Resident Font ROM
      The resident font ROM returned an error during the self-diagnostic testcopyir.com    Replace the controller board
      837    Self-diagnostic Error: Optional Font ROM
      The optional font ROM returned an error during the self-diagnostic testcopyir.com    Replace the controller board
      838    Self-diagnostic Error: Clock Generator
      A verify error occurred when setting data was read from the clock generator via the I2C buscopyir.com    Replace the controller board
      840    EEPROM error 1
      During input/output with the EEPROM, one of the following errors occurred: • A read error occurred and continued after 3 retriescopyir.com • Write error occurredcopyir.com    • EEPROM defective; replace the controller board • EEPROM has reached the end of its service life
      841    EEPROM error 2
      The values read from the three areas during the mirroring check phase did not matchcopyir.com The data is being written into the three areas differentlycopyir.com    • EEPROM defective; replace the controller board • EEPROM has reached the end of its service life
      850    Network I/F Abnormal
      NIB interface errorcopyir.com    • NIB defective • Controller board defective
      851    IEEE 1394 I/F Abnormal    • IEEE1384 interface board defective • Controller board defective
      853    Wireless LAN card error 1
      At startup the wireless LAN board could be accessed, but the wireless LAN card (IEEE 802copyir.com11b or Bluetooth) could not access the boardcopyir.com    • Wireless LAN card not inserted in the wireless LAN board when machine was switched on
      854    Wireless LAN card error 2
      The board that holds the wireless LAN card can be accessed, but the wireless LAN card (802copyir.com11b/Bluetooth) itself cannot be accessed while the machine is operating    • Wireless LAN card has been removed during machine operationcopyir.com
      855    Wireless LAN card error 3
      An error was detected for the wireless LAN card (802copyir.com11b or Bluetooth)copyir.com    • Wireless LAN card defective • Wireless card connection not tight
      856    Wireless LAN board error
      An error is detected for the wireless LAN card (802copyir.com11b or Bluetooth)copyir.com    • Wireless LAN card board defective • PCI connector loose
      857    USB I/F Error
      The USB driver is unstable and generated an errorcopyir.com The USB I/F cannot be usedcopyir.com The USB driver can generate three types of errors: RX, CRC, and STALL errorscopyir.com Only the STALL error can generate this SC codecopyir.com    • USB 2copyir.com0 disconnected • Controller board defective
      860    Startup without HD connection at main power on
      The hard disk connection is not detectedcopyir.com    • Cable between HDC and HDD loose or defective • HDD power connector loose or defective • HDD defective • Replace the controller board
      861    Startup without HD detection at power key on
      The hard disk connection is not detectedcopyir.com    • Cable between HDC and HDD loose or defective • HDD power connector loose or defective • HDD defective Replace the controller board
      862    Maximum number of bad sectors detected on HD
      Up to 101 bad sectors have appeared in the area on the hard disk where image data is archived, and the hard disk may require replacementcopyir.com    SC863 returned while reading data from the HD and the number of registered bad sectors reached 101copyir.com
      863    Startup without HD data lead
      Data stored on the hard disk is not read correctlycopyir.com    A bad sector occurred during operation of the HDD
      864    HD data CRC error
      During operation of the HD, the HD responded with a CRC errorcopyir.com    Data transfer was abnormal in the data read from the HDDcopyir.com
      865    HD access error
      The hard disk detected an errorcopyir.com    Replace the controller board
      870    Address Book Data Error
      Address book data stored on the hard disk was detected as abnormal when it was accessed from either the operation panel or the networkcopyir.com    • Software defective • HDD defective
      871    FCU Flash ROM Error
      The address book written into the flash ROM mounted on the FCU is detected as defectivecopyir.com    • Flash ROM device defective • Replace flash ROM on the MBU
      880    File Format Converter (MLB) Error
      A request for access to the File Format Converter (MLB -Media Link Board) was not answered within the specified timecopyir.com    • MLB defective
      900    Electronic total counter error
      The value of the total counter has already exceeded 9999999    Replace the NVRAM on the controller board
      901    Mechanical total count error
      The SBCU board cannot receive the mechanical total count datacopyir.com    Mechanical total counter defective
      920    Printer Error 1
      An internal application error was detected and operation cannot continuecopyir.com    • Software defective • Insufficient memory
      925    Network File Error
      The file that manages NetFile is corrupted and operation cannot continuecopyir.com    • Software defective • Files on the HDD corrupted
      951    F-gate error at write request
      After the IPU receives an F-gate signal, it receives another F-gate signalcopyir.com    • Download controller firmware • SBCU board defective
      953    Scanner setting error
      The IPU does not respond with the scanner setting signal required to start scanning processingcopyir.com    Download controller firmware
      954    Printer setting error
      The IPU does not respond with the settings that are required to start image processing by the printercopyir.com    • Replace the IPU board • Replace the controller board • Download the controller firmware
      955    Memory setting error
      The IPU does not respond with the settings that are required to start image processing using the memorycopyir.com    • Replace the IPU board • Replace the controller board • Download the controller firmware
      964    Printer ready error
      The printer ready signal is not generated within 17 seconds after the IPU received the print start signalcopyir.com    • Replace the IPU board • Replace the controller board • Download the controller firmware
      984    Print image data transfer error
      The image transfer from the controller to the engine via the PCI bus does not end within 15 s after startingcopyir.com    • Controller board defective • SBCU board defective • Connectors between SBCU and controller loose or defective
      986    Software write parameter setting error
      An unstable area at the storage destination in the settings table is set at NULL for the parameter received by the write modulecopyir.com    • Download controller firmware again
      990    Software performance error
      The software attempted to perform an unexpected operationcopyir.com    • Software defective • Internal parameter incorrect • Insufficient working memory • When this SC occurs, the file name, address, and data will be stored in NVRAMcopyir.com This information can be checked by using SP7-403copyir.com Note the above data and the situation in which this SC occurscopyir.com Then report the data and conditions to your technical control centercopyir.com
      991    Software continuity error
      The software attempted to perform an unexpected operationcopyir.com However, unlike SC990, the object of the error is continuity of the softwarecopyir.com    • No operation requiredcopyir.com This SC code does not appear on the panel, and is only loggedcopyir.com
      992    Unexpected Software Error
      Software encountered an unexpected operation not defined under any SC codecopyir.com    • Software defective • An error undetectable by any other SC code occurred
      995    Machine Type Information Error
      After the machine is powered on, a mismatch is detected between the CPM information sent from the controller to the enginecopyir.com The controller boards of the B089 (22 cpm) and B94 (27 cpm) are not interchangeablecopyir.com    • Replace the controller board with the correct boardcopyir.com
      996    FCU board error
      FCU board is connected but not readycopyir.com    • FCU board defective and requires replacement • Download FCU firmware
      997    Application function selection error
      The application selected by a key press on operation panel does not start or ends abnormallycopyir.com    • Download the firmware for the application that failed • An option required by the application (RAM, DIMM, board) is not installed
      998    Application start error
      After power on, the application does not start within 60 scopyir.com (All applications neither start nor end normallycopyir.com)    • Download controller firmware • Replace the controller board • An option required by the application (RAM, DIMM, board) is not installed
      999    Program download error
      The download (program, print data, language data) from the IC card does not execute normallycopyir.com    • Board installed incorrectly • SBCU board defective • Controller board defective • IC card defective • NVRAM defective • Loss of power during downloading • Important Notes About SC999 • Primarily intended for operating in the download mode, logging is not performed with SC999copyir.com • If the machine loses power while downloading, or if for some other reason the download does not end normally, this could damage the controller board or the PCB targeted for the download and prevent subsequent downloadingcopyir.com If this problem occurs, the damaged PCB must be replacedcopyir.com

101    Exposure lamp error
The standard white level was not detected properly when scanning the white platecopyir.com    • Exposure lamp defective • Lamp stabilizer defective • Exposure lamp connector defective • Standard white plate dirty • Scanner mirror or scanner lens out of position or dirty • SBU defective
120    Scanner home position error 1
The scanner home position sensor does not detect the on condition during initialization or copyingcopyir.com    • Scanner drive motor defective • Scanner motor defective • Harness between SBCU and scanner drive motor disconnected • Harness between SBCU and scanner drive motor power source disconnected • Scanner HP sensor defective • Harness between SBCU and HP sensor disconnected • Scanner wire, timing belt, pulley, or carriage defective
121    Scanner home position error 2
The scanner home position sensor does not detect the off condition during initializationcopyir.com    • SIB or scanner drive motor defective • Scanner motor defective • Harness between SBCU and scanner drive motor disconnected • Harness between SBCU and scanner drive motor power source disconnected • Scanner HP sensor defective • Harness between SBCU and HP sensor disconnected • Scanner wire, timing belt, pulley, or carriage defective
122    Scanner HP Sensor œ Error 1
The HP sensor remains on while the carriage is returning to the home positioncopyir.com    • SBCU, scanner motor drive board defective • Scanner motor defective • Harness between the SBCU, scanner motor drive board and scanner motor disconnected • HP sensor defective • Harness between SBCU and HP sensor disconnectedcopyir.com • Scanner wire, timing belt, pulley, or carriage installation incorrect
123    Scanner HP sensor œ Error 2
The HP sensor does not switch on after the carriage has returned to the home positioncopyir.com    • SBCU, scanner motor drive board defective • Scanner motor defective • Harness between the SBCU, scanner motor drive board and scanner motor disconnected • HP sensor defective • Harness between SBCU and HP sensor disconnectedcopyir.com • Scanner wire, timing belt, pulley, or carriage installation incorrect
144    SBU connection error
The IPU does not detect the SBU connection signalcopyir.com    Cable between SBU and IPU board disconnectedcopyir.com
192    Automatic SBU adjustment error
An error is detected during automatic SBU adjustment (SP4-428)    • SBU defective • SBCU board defective • Exposure lamp stabilizer defective • Exposure lamp defective • Dirty white plate
193    Image transfer error
The IPU board does not finish within 1 minute after the image data has been transferred to the controller boardcopyir.com    • IPU board defective • Controller board defective • Video controller defective
195    DFGATE assert error
The DFGATE signal does not assert within 30 seconds after the original has been scannedcopyir.com    • ADF interface cable defective • SBCU board defective • Mismatched firmware between the SBCU board and ADF
196    DFGATE negate error
The DFGATE signal does not negate within 1 minute after the DFGATE has been assertedcopyir.com    • ADF interface cable defective • SBCU board defective • Mismatched firmware between the SBCU board and ADF
197    DFGATE error
The DFGEATE signal has already been asserted at the original scancopyir.com    • ADF interface cable defective • SBCU board defective • Mismatched firmware between the SBCU board and ADF
198    Memory address error
The IPU board does not receive the memory address from the controller boardcopyir.com    • Mismatched firmware between the SBCU board and controller board • Controller defective • SBCU defective • IPU board defective
199    DF scanning finish error
The original does not finish scanning with in 1 minute    • ADF interface cable defective • SBCU board defective • Mismatched firmware between the SBCU board and ADF
302    Charge roller current leak
A charge roller current leak signal is detectedcopyir.com    • Charge roller damaged • Charge high voltage supply board defective • Poor connection of the PCU
320    Polygon motor error
The polygon motor does not reach its operating speed within 10 seconds after the polygon motor on signal, or the lock signal is not detected for more than 200 ms continuously during operationcopyir.com    • Polygon mirror motor defective • Poor connection between the polygon mirror motor driver and the SBCU board • SBCU board defective
322    Laser synchronization error
The main scan synchronization detector board cannot detect the laser synchronization signal for more than 10 consecutive 50 ms intervalscopyir.com    • Poor connection between the laser synchronization detector board and the SBCU board • Laser synchronization detector board out of position • Laser synchronization detector board defective • SBCU board defective • LD unit defective
323    LD drive current over
The LD drive board applies more than 100 mA to the LDcopyir.com    • LD unit defective (not enough power, due to aging) • Poor connection between the LD unit and the SBCU board • SBCU board defective
350    ID sensor calibration - Error 1
One of the following conditions occurred when the ID sensor pattern was calibrated during printing: • Vsp > 2copyir.com5V • Vsg < 2copyir.com5V • Vsp = 0V • Vsg = 0V    • ID sensor defective or dirty • ID sensor harness disconnected or connector is damaged • SBCU defective • Scanning system or image creation system malfunction • High voltage power supply board (power pack) defective
351    ID sensor calibration œ Error 2
The following conditions occurred simultaneously when the ID sensor pattern was calibrated during printing: • Vsg = 5V • PWM = 0 (LED current drop)    • ID sensor dirty or defective • ID sensor harness disconnected, or connector damaged • SBCU board defective • High voltage power supply board (power pack) defective
352    ID sensor calibration œ Error 3
During printing the 2copyir.com5V value for edge detection of the ID sensor pattern could not be detected after 800 mscopyir.com    • ID sensor dirty or defective • ID sensor harness disconnected, or connector damaged • SBCU defective • High voltage power supply board (power pack) defective
353    ID sensor adjustment Error 1
Error occurred during automatic adjustment of Vsg: • Vsg output did not attain 4V, even with PWM = 255 (maximum current for LED) • Vsg output was greater than 4V, even with PWM=0 (no current for the LED)    • ID sensor dirty or defective • ID sensor harness disconnected, or connector damaged • SBCU defective • High voltage power supply board (power pack) defective • Scanning system or image creation system malfunction
354    ID Sensor Adjustment Error 2
Error occurred during automatic adjustment of Vsgcopyir.com Vsg could not be adjusted to 4copyir.com0V±0copyir.com2V within 50 ms even after 20 attemptscopyir.com    • ID sensor dirty or defective • ID sensor harness disconnected, or connector damaged • SBCU defective • High voltage power supply board (power pack) defective • Scanning system or image creation system malfunction
355    ID sensor error
For details about the cause of the problem, please refer to SC350~354 abovecopyir.com    • ID sensor dirty or defective • ID sensor harness disconnected, or connector damaged • SBCU board defective • High voltage power supply board (power pack) defective • Scanning system or image creation system malfunction
389    TD sensor error
TD sensor output was less than 0copyir.com5V, or more than 0copyir.com5V 10 times in successioncopyir.com If the fax unit is installed, this SC is issued immediatelycopyir.com If the fax unit is not installed, this SC is issued after the prescribed number of copies has printedcopyir.com    • TD sensor defective • TD sensor connector damagedcopyir.com
390    TD sensor error
The TD sensor outputs less than 0copyir.com5V or more than 4copyir.com0V 10 times consecutively during copyingcopyir.com    • TD sensor abnormal • Poor connection of the PCU
391    Development bias leak
A development bias leak signal is detectedcopyir.com    • Poor connection of the PCU • High voltage supply board defective
392    TD sensor initial setting error
TD sensor output voltage falls out of the adjustment range (2copyir.com0 ±<> 0copyir.com2 V) after the TD sensor initial setting has been finishedcopyir.com    • Someone forgot to remove the toner seal of the PCU • ID sensor defective • TD sensor abnormal • Drum does not turn • Development roller does not turn • Poor connection of the PCU
399    Illegal toner bottle (South Korea only)
The toner bottle installed is not intended for use with this machinecopyir.com    • Install the correct type of toner bottlecopyir.com
401    Transfer roller leak error 1
A transfer roller current leak signal is detectedcopyir.com The current feedback signal for the transfer roller is not detectedcopyir.com    • High voltage supply board defective • Poor connection of the PCU • Transfer/separation unit set incorrectly • Transfer roller damaged
402    Transfer roller leak error 2
A transfer roller current leak signal is detectedcopyir.com The current feedback signal for the transfer roller is not detectedcopyir.com    • High voltage supply board defective • Poor connection of the PCU • Transfer/separation unit set incorrectly • Transfer roller damaged
411    Separation bias leak error
A separation bias leak signal is detectedcopyir.com    • High voltage supply board defective • Poor connection of the PCU • Discharge plate defective
490    Toner supply motor leak error
Over 1 A supplied to the toner supply motor for longer than 200 ms    Toner supply motor defective
500    Main motor lock
A main motor lock signal is not detected for more than 500 ms after the main motor starts to rotate, or the lock signal is not detected for more than 500 ms during rotation after the last signalcopyir.com    • Too much load on the drive mechanism • Main motor defective
501    1st paper tray lift motor malfunction    • Paper lift sensor defective • Tray lift motor defective • Too much load on the drive mechanism • Poor tray lift motor connection
502    2nd paper tray lift motor malfunction    • Paper lift sensor defective • Tray lift motor defective • Too much load on the drive mechanism • Poor tray lift motor connection
503    3rd paper tray lift motor malfunction (optional paper tray unit)    • Paper lift sensor defective • Tray lift motor defective • Too much load on the drive mechanism • Poor tray lift motor connection
504    4th paper tray lift motor malfunction (optional paper tray unit)    • Paper lift sensor defective • Tray lift motor defective • Too much load on the drive mechanism • Poor tray lift motor connection
506    Paper tray motor lock (optional paper tray unit)
A motor lock signal is not detected for more than 1copyir.com5 s or the lock signal is not detected for more than 1copyir.com0 s during rotationcopyir.com    • Paper tray unit motor defective • Too much load on the drive mechanism
508    Rear fence drive error (optional LCT)
The return position sensor is not activated after the rear fence drive motor has been on to lower the tandem tray for 8 secondscopyir.com    • Rear fence motor defective • Return position sensor defective • Too much load on the drive mechanism
509    Side fence drive error (optional LCT)
The side fence positioning sensor is not activated for more 3 seconds when the paper stack in the left tray is moved to the right traycopyir.com The side fence close sensor is not activated for more 3 seconds after moving the paper stack to the right traycopyir.com    • Side fence motor defective • Side fence position sensor defective • Side fence close sensor defective • Too much load on the drive mechanism
510    LCT lower limit error
The lower limit sensor does not activate within 8 seconds after the tray has been loweredcopyir.com    • Tray lift motor defective • Poor connection of the tray lift motor • Lower limit sensor defective • Too much load on the drive mechanism
520    Paper tray error
An error occurs (icopyir.come motor error, or sensor error, etc) for any paper traycopyir.com    • A defective motor • A defective sensor • Too much load on the drive mechanism
541    Fusing thermistor open
The fusing temperature detected by the thermistor was below 0 °C for 5 secondscopyir.com The fusing temperature does not rise +15 °C (center) or +12 °C (at the ends) five times within 2 minutes after the fusing lamps have been turned oncopyir.com    • Fusing thermistor defective or out of position • Fusing lamp open • Fusing thermostat open • Power supply board defective • Poor connection of the fusing unit
542    Fusing temperature warm-up error
The fusing temperature does not reach the standby temperature within 20 seconds after the main switch is turned oncopyir.com    • Fusing thermistor defective or out of position • Fusing lamp open • Fusing thermostat open • Power supply board defective • Poor connection of the fusing unit
543    Fusing overheat error 1
A fusing temperature of over 231°C is detected for 5 second by the fusing thermistorcopyir.com    • Fusing thermistor defective • Power supply board defective • SBCU board defective
544    Fusing overheat error 2
A fusing temperature of over 251°C is detected by the fusing temperature monitor circuit in the SBCU boardcopyir.com    • Fusing thermistor defective • Power supply board defective • SBCU board defective
545    Fusing overheat error 3
After warmup, the hot roller attained full operating temperature and maintained this temperature for 10 seccopyir.com without the hot roller rotatingcopyir.com    • Hot roller thermistor is out of position • Fusing lamp broken • Thermostat broken
546    Unstable fusing temperature
The fusing temperature varies 50°C or more within 1 second twice continuouslycopyir.com    • Thermistor defective • Poor connection of the fusing unit • Power supply unit defective
547    Zero cross signal malfunction
Zero cross signals are not detected within a certain periodcopyir.com    • Power supply board defective • SBCU board defective
548    Fusing unit set error
The machine does not detect the fusing unitcopyir.com    • Poor connection of the fusing unit • The fusing unit is not installed
590    Exhaust fan motor error
The CPU detects an exhaust fan lock signal for more than 5 secondscopyir.com    • Poor connection of the exhaust fan motor • Too much load on the motor drive
611    Communication break error between SBCU and ADF
The SBCU receives a break signal from the ADF main boardcopyir.com    • Serial line connecting SBCU and ADF unstable • External noise • ADF main board defective • SBCU board defective
612    Communication command error between SBCU and ADF
The SBCU sends a command to the ADF main board that it cannot executecopyir.com    • SBCU board defective • Download SBCU firmware again
620    Communication timeout error between SBCU and finisher
The SBCU cannot receive a response within 100 ms after 3 attempts after sending data to the finishercopyir.com    • Serial line connecting SBCU and finisher unstable • External noise • SBCU board and finisher main board connection defective or loose • Finisher main board defective • SBCU board defective
621    Communication timeout error between SBCU and finisher
A break (low) signal was received from the finishercopyir.com    • Serial line connecting SBCU and finisher unstable • External noise
630    Communication failure with CSS (RSS)
The communication from the copier was detected as abnormal at the CSS centercopyir.com This error occurs when the acknowledge signal from the LADP does not complete normallycopyir.com    • Occurred with an SC call, CC call, supply management call, user call, or CE callcopyir.com • Timeout while no response from the LADP, and abnormal signal on the RS-485 line between PI and LADPcopyir.com
632    Counter device error 1
After 3 attempts to send a data frame to the optional counter device via the serial communication line, no ACK signal was received within 100 mscopyir.com    • Serial line between the optional counter device, the relay board and copier control board is disconnected or damaged • Make sure that SP5113 is set to enable the optional counter devicecopyir.com
633    Counter device error 2
After 3 attempts to send a data frame to the optional counter device via the serial communication line, no ACK signal was received within 100 mscopyir.com    • Serial line between the optional counter device, the relay board and copier control board is disconnected or damaged • Make sure that SP5113 is set to enable the optional counter devicecopyir.com
634    Counter device error 3
A backup RAM error was returned by the counter devicecopyir.com    • Counter device control board defective • Backup battery of counter device defective
635    Counter device error 4
A backup battery error was returned by the counter devicecopyir.com    • Counter device control board defective • Backup battery of counter device defective
640    SBCU control data transfer checksum error
A sampling of control data sent from the SBCU to the controller reveals a checksum errorcopyir.com Only the logging count is performedcopyir.com    • Controller board defective • External noise • SBCU board defective
641    SBCU control data transfer abnormal
A sampling of the control data sent from the SBCU reveals an abnormalitycopyir.com    • Controller board defective • External noise • SBCU board defective
650    Communication timeout error between SBCU and duplex unit
The SBCU cannot receive a response within 1 second from the duplex unitcopyir.com    • Serial line connecting SBCU and duplex unit unstable • External noise • SBCU board and duplex main board connection defective or loose • Duplex main board defective • SBCU board defective
670    Engine response error
After powering on the machine, a response is not received from the engine within 30 secondscopyir.com    • SBCU installed incorrectly • SBCU defective • Controller board defective
672    Controller-to-operation panel communication error at startup
After powering on the machine, the communication circuit between the controller and the operation panel is not opened, or communication with controller is interrupted after a normal startupcopyir.com    • Controller stall • Controller board installed incorrectly • Controller board defective • Operation panel connector loose or defective
720    Finisher jogger motor error (500-sheet finisher)
The finisher jogger Hcopyir.comP sensor remains de-activated for a certain time when returning to home positioncopyir.com The finisher jogger Hcopyir.comP sensor remains activated for a certain time when moving away from home positioncopyir.com    • Jogger Hcopyir.comP sensor defective • Jogger motor defective
722    Finisher jogger motor error (1000-sheet finisher)
The finisher jogger Hcopyir.comP sensor remains de-activated for a certain time when returning to home positioncopyir.com The finisher jogger Hcopyir.comP sensor remains activated for a certain time when moving away from home positioncopyir.com    • Jogger Hcopyir.comP sensor defective • Jogger motor defective
724    Finisher staple hammer motor error (1000-sheet finisher)
Stapling does not finish within 600 ms after the staple hammer motor turned oncopyir.com    • Staple jam • Stapler overload caused by trying to staple too many sheets • Staple hammer motor defective
725    Finisher stack feed-out motor error (1000-sheet finisher)
The stack feed-out belt Hcopyir.comP sensor does not activate within a certain time after the stack feed-out motor turned oncopyir.com    • Stack feed-out Hcopyir.comP sensor defective • Stack feed-out motor defective
726    Finisher lift motor error (1000-sheet finisher)
The stack height sensor does not activate within a certain time after the shift tray lift motor turned oncopyir.com    • Shift tray lift motor defective • Stack height sensor defective
727    Finisher staple hammer motor error (500-sheet finisher)
Stapling does not finish within a certain time after staple hammer motor turned oncopyir.com    • Staple jam • Stapler overload caused by trying to staple too many sheets • Staple hammer motor defective
728    Finisher paper stack height error (500sheet finisher)
The stack height detection lever does not return to its home position before going to detect the stack heightcopyir.com    • Stack height lever solenoid defective • Stack height sensor defective • Lever sensor defective • Main control board defective
730    Finisher stapler motor error (1000-sheet finisher)
The stapler does not return to its home position within a certain time after the stapler motor turned oncopyir.com The stapler Hcopyir.comP sensor does not activate within a certain time after the stapler motor turned oncopyir.com    • Stapler motor defective • Stapler Hcopyir.comP sensor defective • Poor stapler motor connection
731    Output tray motor error (500-sheet finisher) Exit guide plate motor error (1000-sheet finisher)
The tray upper limit sensor does not activate within a certain time after the shift motor turned oncopyir.com (500-sheet finisher) The exit guide plate open sensor or exit guide plate HP sensor does not activate within a certain time after the exit guide plate motor turned oncopyir.com (1000-sheet finisher)    500-sheet finisher • Output tray motor defective • Tray upper limit sensor defective 1000-sheet finisher • Exit guide plate motor defective • Exit guide plate HP sensor defective • Exit guide plate open sensor defective
732    Finisher shift motor error (1000-sheet finisher)
Roller shift does not finish within a certain time after the shift motor turned oncopyir.com    • Shift motor defective • Shift tray HP sensor defective
770    Shift tray shift motor error
During a shift operation the sensor state did not change (off to on, or on to off) within 2copyir.com4 seccopyir.com    • Shift sensor defective • Shift motor defective
791    Bridge communication error
The SBCU cannot communicate with the bridge unit properly when the finisher is installedcopyir.com    • Poor connection between the main machine and bridge unit • SBCU board defective
792    Finisher connection error
The SBCU cannot communicate with the finisher properly when the bridge unit is installedcopyir.com    • Poor connection between the finisher and bridge unit • SBCU board defective
793    Interchange communication error
The SBCU cannot communicate with the interchange unit properly when the duplex unit is installedcopyir.com    • Poor connection between the interchange unit and main machine • SBCU board defective
800    Startup without video output end error (K)
Video transfer to the engine is started, but the engine did not issue a video transmission end command within the specified timecopyir.com    Controller board defective
804    Startup without video input end (K)
A video transmission was requested from the scanner, but the scanner did not issue a video transmission end command within the specified timecopyir.com    Controller board defective
818    Watchdog error
The CPU does not access the watchdog register within a certain timecopyir.com    • Controller board defective • Software malfunction œ download controller firmware again
819    Kernel mismatch error
Software bug    Download controller firmware again
820    Self-Diagnostic Error: CPU
The central processing unit returned an error during the self-diagnostic testcopyir.com    • Controller board defective • Download controller firmware again
821    Self-Diagnostic Error: ASIC
The ASIC returned an error during the self-diagnostic test because the ASIC and CPU timer interrupts were compared and determined to be out of rangecopyir.com    Controller board defective
822    Self-Diagnostic Error: HDD
The hard disk drive returned an error during the self-diagnostic testcopyir.com    • HDD defective • HDD connector defective • Controller board defective
823    Self-diagnostic Error: NIB
The network interface board returned an error during the self-diagnostic testcopyir.com    • Network interface board defective • Controller board defective
824    Self-diagnostic Error: Resident NVRAM
The resident non-volatile RAM returned an error during the self-diagnostic testcopyir.com    • Replace the resident NVRAM on the controller board • Replace the controller board
825    Self-diagnostic Error: Optional NVRAM
The resident non-volatile RAM returned an error during the self-diagnostic testcopyir.com    • Replace the optional NVRAM (user account enhancement unit) on the controller board • Replace the controller board
826    Self-diagnostic Error: NVRAM/Optional NVRAM
The NVRAM or optional NVRAM returned an error during the self-diagnostic testcopyir.com    Replace the NVRAM on the controller board
827    Self-diagnostic Error: RAM
The resident RAM returned a verify error during the self-diagnostic testcopyir.com    Download controller firmware again
828    Self-diagnostic Error: ROM
The resident read-only memory returned an error during the self-diagnostic testcopyir.com    • Controller board defective • Download controller firmware again
829    Self-diagnostic Error: Optional RAM
The optional RAM returned an error during the self-diagnostic testcopyir.com    • Replace the optional memory board • Controller board defective
835    Self-Diagnostic Error: Parallel Interface
Loopback test errorcopyir.com    • Loopback connector not detected • IEEE1284 connector defective • Controller board defective
836    Self-diagnostic Error: Resident Font ROM
The resident font ROM returned an error during the self-diagnostic testcopyir.com    Replace the controller board
837    Self-diagnostic Error: Optional Font ROM
The optional font ROM returned an error during the self-diagnostic testcopyir.com    Replace the controller board
838    Self-diagnostic Error: Clock Generator
A verify error occurred when setting data was read from the clock generator via the I2C buscopyir.com    Replace the controller board
840    EEPROM error 1
During input/output with the EEPROM, one of the following errors occurred: • A read error occurred and continued after 3 retriescopyir.com • Write error occurredcopyir.com    • EEPROM defective; replace the controller board • EEPROM has reached the end of its service life
841    EEPROM error 2
The values read from the three areas during the mirroring check phase did not matchcopyir.com The data is being written into the three areas differentlycopyir.com    • EEPROM defective; replace the controller board • EEPROM has reached the end of its service life
850    Network I/F Abnormal
NIB interface errorcopyir.com    • NIB defective • Controller board defective
851    IEEE 1394 I/F Abnormal    • IEEE1384 interface board defective • Controller board defective
853    Wireless LAN card error 1
At startup the wireless LAN board could be accessed, but the wireless LAN card (IEEE 802copyir.com11b or Bluetooth) could not access the boardcopyir.com    • Wireless LAN card not inserted in the wireless LAN board when machine was switched on
854    Wireless LAN card error 2
The board that holds the wireless LAN card can be accessed, but the wireless LAN card (802copyir.com11b/Bluetooth) itself cannot be accessed while the machine is operating    • Wireless LAN card has been removed during machine operationcopyir.com
855    Wireless LAN card error 3
An error was detected for the wireless LAN card (802copyir.com11b or Bluetooth)copyir.com    • Wireless LAN card defective • Wireless card connection not tight
856    Wireless LAN board error
An error is detected for the wireless LAN card (802copyir.com11b or Bluetooth)copyir.com    • Wireless LAN card board defective • PCI connector loose
857    USB I/F Error
The USB driver is unstable and generated an errorcopyir.com The USB I/F cannot be usedcopyir.com The USB driver can generate three types of errors: RX, CRC, and STALL errorscopyir.com Only the STALL error can generate this SC codecopyir.com    • USB 2copyir.com0 disconnected • Controller board defective
860    Startup without HD connection at main power on
The hard disk connection is not detectedcopyir.com    • Cable between HDC and HDD loose or defective • HDD power connector loose or defective • HDD defective • Replace the controller board
861    Startup without HD detection at power key on
The hard disk connection is not detectedcopyir.com    • Cable between HDC and HDD loose or defective • HDD power connector loose or defective • HDD defective Replace the controller board
862    Maximum number of bad sectors detected on HD
Up to 101 bad sectors have appeared in the area on the hard disk where image data is archived, and the hard disk may require replacementcopyir.com    SC863 returned while reading data from the HD and the number of registered bad sectors reached 101copyir.com
863    Startup without HD data lead
Data stored on the hard disk is not read correctlycopyir.com    A bad sector occurred during operation of the HDD
864    HD data CRC error
During operation of the HD, the HD responded with a CRC errorcopyir.com    Data transfer was abnormal in the data read from the HDDcopyir.com
865    HD access error
The hard disk detected an errorcopyir.com    Replace the controller board
870    Address Book Data Error
Address book data stored on the hard disk was detected as abnormal when it was accessed from either the operation panel or the networkcopyir.com    • Software defective • HDD defective
871    FCU Flash ROM Error
The address book written into the flash ROM mounted on the FCU is detected as defectivecopyir.com    • Flash ROM device defective • Replace flash ROM on the MBU
880    File Format Converter (MLB) Error
A request for access to the File Format Converter (MLB -Media Link Board) was not answered within the specified timecopyir.com    • MLB defective
900    Electronic total counter error
The value of the total counter has already exceeded 9999999    Replace the NVRAM on the controller board
901    Mechanical total count error
The SBCU board cannot receive the mechanical total count datacopyir.com    Mechanical total counter defective
920    Printer Error 1
An internal application error was detected and operation cannot continuecopyir.com    • Software defective • Insufficient memory
925    Network File Error
The file that manages NetFile is corrupted and operation cannot continuecopyir.com    • Software defective • Files on the HDD corrupted
951    F-gate error at write request
After the IPU receives an F-gate signal, it receives another F-gate signalcopyir.com    • Download controller firmware • SBCU board defective
953    Scanner setting error
The IPU does not respond with the scanner setting signal required to start scanning processingcopyir.com    Download controller firmware
954    Printer setting error
The IPU does not respond with the settings that are required to start image processing by the printercopyir.com    • Replace the IPU board • Replace the controller board • Download the controller firmware
955    Memory setting error
The IPU does not respond with the settings that are required to start image processing using the memorycopyir.com    • Replace the IPU board • Replace the controller board • Download the controller firmware
964    Printer ready error
The printer ready signal is not generated within 17 seconds after the IPU received the print start signalcopyir.com    • Replace the IPU board • Replace the controller board • Download the controller firmware
984    Print image data transfer error
The image transfer from the controller to the engine via the PCI bus does not end within 15 s after startingcopyir.com    • Controller board defective • SBCU board defective • Connectors between SBCU and controller loose or defective
986    Software write parameter setting error
An unstable area at the storage destination in the settings table is set at NULL for the parameter received by the write modulecopyir.com    • Download controller firmware again
990    Software performance error
The software attempted to perform an unexpected operationcopyir.com    • Software defective • Internal parameter incorrect • Insufficient working memory • When this SC occurs, the file name, address, and data will be stored in NVRAMcopyir.com This information can be checked by using SP7-403copyir.com Note the above data and the situation in which this SC occurscopyir.com Then report the data and conditions to your technical control centercopyir.com
991    Software continuity error
The software attempted to perform an unexpected operationcopyir.com However, unlike SC990, the object of the error is continuity of the softwarecopyir.com    • No operation requiredcopyir.com This SC code does not appear on the panel, and is only loggedcopyir.com
992    Unexpected Software Error
Software encountered an unexpected operation not defined under any SC codecopyir.com    • Software defective • An error undetectable by any other SC code occurred
995    Machine Type Information Error
After the machine is powered on, a mismatch is detected between the CPM information sent from the controller to the enginecopyir.com The controller boards of the B089 (22 cpm) and B94 (27 cpm) are not interchangeablecopyir.com    • Replace the controller board with the correct boardcopyir.com
996    FCU board error
FCU board is connected but not readycopyir.com    • FCU board defective and requires replacement • Download FCU firmware
997    Application function selection error
The application selected by a key press on operation panel does not start or ends abnormallycopyir.com    • Download the firmware for the application that failed • An option required by the application (RAM, DIMM, board) is not installed
998    Application start error
After power on, the application does not start within 60 scopyir.com (All applications neither start nor end normallycopyir.com)    • Download controller firmware • Replace the controller board • An option required by the application (RAM, DIMM, board) is not installed
999    Program download error
The download (program, print data, language data) from the IC card does not execute normallycopyir.com    • Board installed incorrectly • SBCU board defective • Controller board defective • IC card defective • NVRAM defective • Loss of power during downloading • Important Notes About SC999 • Primarily intended for operating in the download mode, logging is not performed with SC999copyir.com • If the machine loses power while downloading, or if for some other reason the download does not end normally, this could damage the controller board or the PCB targeted for the download and prevent subsequent downloadingcopyir.com If this problem occurs, the damaged PCB must be replacedcopyir.com

انواع دستگاه های فتوکپی لطفا کلیک کنید

انواع دستگاهای فتوکپی رومیزی  لطفا کلیک کنید 

انواع دستگاه های فتوکپی ایستاده لطفا کلیک کنید 

انواع دستگاهای فتوکپی تکثیر انبوه لطفا کلیک کنید 

انواع دستگاهای ریسوکراف RZ لطفا کلیک کنید 

انواع دستگاهای ریسوگراف MZ لطفا کلیک کنید

انواع دستگاهای ریسوگراف RP لطفا کلیک کنید

انواع دستگاهای ریسوگراف EZ لطفا کلیک کنید 

انواع دستگاهای ریسو گرافهای رنگی لطفا کلیک کنید 

انواع دستگاهای پلاتر های رنگی  لطفا کلیک کنید

انواع دستگاهای لمینت لطفا کلیک کنید 

انواع کاترهای دستی و برقی لطف کلیک کنید

انواع دستگاهی صحافی لطفا کلیک کنید 

انواع دوخت لطفا کلیک کنید 

انواع تونر و مواد مصرفی لطفا کلیک کنید 

 

خواندن 1081 دفعه آخرین ویرایش در پنج شنبه, 06 تیر 1398 11:44

نظر دادن

از پر شدن تمامی موارد الزامی ستاره‌دار (*) اطمینان حاصل کنید. کد HTML مجاز نیست.

نماد اعتماد

اطلاعات تماس

 آدرس: آذربایجان شرقی، مراغه، خیایان امام، روبروی داروخانه جاوید، ساختمان کپی ایران
شماره تماس: 04137252392   |    04137254200
شماره تماس: 04137253033   |   04137253034
شماره همراه: 09148659399
شماره فکس: 04136586892
پست الکترونیکی: این آدرس ایمیل توسط spambots حفاظت می شود. برای دیدن شما نیاز به جاوا اسکریپت دارید.این آدرس ایمیل توسط spambots حفاظت می شود. برای دیدن شما نیاز به جاوا اسکریپت دارید|این آدرس ایمیل توسط spambots حفاظت می شود. برای دیدن شما نیاز به جاوا اسکریپت داریدساعات کاری: از 8 الی 22 شب